    Heat Deflection Temperature and Vicat Tester(HDT/VICAT)

    With reliable quality and advanced technology,
    we deliver solutions optimized to meet a wide range of customer needs.

    Heat Deflection Temperature and Vicat Tester(HDT/VICAT)
    • PDF Download
    • Product overview

      Thermal deformation softening point tester is an instrument that measures the thermal deformation of hard plastics and the temperature of non-cart softening point of thermoplastic according to constant load and temperature changes.

      Product Features

      Testing by international standards other than ISO 75, ISO 306, ASTM D648, and ASTM 1526

      Fully automated test cycle support

      Installation of air bearing guide method to prevent friction of load piston

      Precisely measuring bending deformation and press-fitting sides up to 0.001 mm (0.001 in) using an electric conversion device

      Water-cooled circulation (inside heat exchanger) supports fast temperature drop rate after completion of the test

      Use air to support the test station's one-touch up-down function (test piece, easy to attach and detach)

    Product specifications

    Model Unit 603 303
    Number of Stations 6 3
    Temperature Range 23° to 300°
    Temperature Ramp 50per hour of 2per minute
    Temperature Display Resolution 0.1
    Temperature Sensor Platinum RTD located adjacent to the load application point at each station
    Deflection/Penetration Measurement LVDT
    Deflection/Penetration Display Resolution mm 0.001
    Cooldown Rate Maximm of 20 above cooling water temperature in 20minutes
    Temperature Safety Limit independent dual systems using thermostatic switch in bath and keypad selectable software limiting
    Dimensions
    (W×D×H)
    mm 969×769×572 826×639×585
    in 38.5×30.5×22.5 32.5×25.5×23
    Weight kg 132 87
    lb 290 190
    Required Utilities Heat Transfer Medium 16liters (3.7 US gallons) min 11liters (2.9 US gallons)
    Water Water supply for cooldown
    Clean Air Dry air filtered to 50 microns at 40psi (2.7bar) min
    Power 220 ± 10%, 50/60Hz, 1phase, 4.5kW